Specifications
Maximum voltage between any terminal and earth ground
600 V
Surge protection
6 kV peak per IEC 61010–1 600 VCAT III, Pollution Degree 2
Display
Digital: 6,000 counts, updates 4 per second
Bar graph
33 segments, updates 32 per second
Operating temperature
-10°C to + 50°C
Storage temperature
-40°C to + 60°C
Battery type
9 volt Alkaline, NEDA 1604A/ IEC 6LR61
Battery life
400 hours typical, without backlight
Accuracy Specifications
DC millivolts
Range/resolution
600.0 mV / 0.1 mV
Accuracy
±([% of reading] + [counts]): 0.5% + 2
DC volts
Range/resolution
6.000 V / 0.001 V
60.00 V / 0.01 V
600.00 V / 0.1 V
Accuracy
±([% of reading] + [counts]): 0.5% + 2
Auto volts
Range/resolution
600.0 V / 0.1 V
Accuracy
2.0% + 3 (DC, 45 Hz to 500 Hz)
4.0% + 3 (500 Hz to 1 kHz)
AC millivolts1 True RMS
Range/resolution
600.0 mV / 0.1 mV
Accuracy
1.0% + 3 (DC, 45 Hz to 500 Hz)
2.0% + 3 (500 Hz to 1 kHz)
AC volts1 True RMS
Range/resolution
6.000 V / 0.001 V
60.00 V / 0.01 V
600.0 V / 0.1 V
Accuracy
1.0% + 3 (DC, 45 Hz to 500 Hz)
2.0% + 3 (500 Hz to 1 kHz)
Continuity
Range/resolution
600 Ω / 1 Ω
Accuracy
Beeper on < 20 Ω, off > 250 Ω detects opens or shorts of
500 μs or longer.
Ohms
Range/resolution
600.0 Ω / 0.1 Ω
6.000 kΩ / 0.001 kΩ
60.00 kΩ / 0.01 kΩ
600.0 kΩ / 0.1 kΩ
6.000 MΩ / 0.001 MΩ
Accuracy
0.9% + 1
Range/Resolution
40.00 MΩ / 0.01 MΩ
Accuracy
5% + 2
Diode Test
Range/resolution
2.00 V / 0.001 V
Accuracy
0.9% + 2
Capacitance
Range/resolution
1000 nF / 1 nF
10.00 μF / 0.01 μF
100.0 μF / 0.1 μF
9999 μF / 1 μF
100 μF to 1000 μF
Accuracy
1.9% + 2
Range/resolution
> 1000 μF
Accuracy
5% + 20%
Lo-Z Capacitance
Range
1 nF to 500 μF
Accuracy
10% + 2 typical
Temperature²
(Type-K thermocouple)
Range/resolution
-40°C to 400°C / 0.1°C
Accuracy
1% + 10²
Range/resolution
-40°F to 752°F / 0.2°F
Accuracy
1% + 18²
AC μA true-rms (45 Hz to 500 Hz)
Range/resolution
600.0 μA / 0.1 μA
Accuracy
1.5% + 3
DC μamps
Range/resolution
600.0 μA / 0.1 μA
Accuracy
1.0% + 2
Hz (V or A input)²
Range/resolution
99.99 Hz / 0.01 Hz
999.99 Hz / 0.1 Hz
9.999 Hz / 0.001 Hz
50.00 Hz / 0.01 Hz
Accuracy
0.1% + 2
1. All AC voltage ranges except Auto-V/LoZ are specified from 1% to 100% of range. Auto-V/LoZ is specified from 0.0 V.
2. Temperature uncertainty (accuracy) does not include the error of the thermocouple probe.
3. Frequency is AC coupled, 5 Hz to 50 kHz for AC voltage. Frequency is DC coupled, 45 Hz to 5 kHz for AC current.
4. Temperature Range/Resolution: -40°F to 752°F / 0.2°F.
Mechanical and General Specifications
Size
167 x 84 x 46 mm (6.57 x 3.31 x 1.82 in)
Weight
550 g
Warranty